Looking forward to your cooperation with us

By 2025, we will work hand in hand with you for a win-win situation, providing your customers with high-quality products and efficient technical services!

Home >> Product Center >> Research Products >> Human
wechat_2025-10-14_143837_721.png

C-peptide ELISA

Product Item Number:10-1136-01
Product Specifications:1 x 96 wells
Product Remarks:
Measurement range: 100 - 4000 pmol/L (0.302 - 12.1 µg/L)Featureü No or minimal cross-reactivity to human insulin and proinsulinü High specificity enables measurement of human C-peptide in mouse and rat modelsü CE/IVD labeled on selected marketsü Serum controls available:
Product Description

Measurement range: 100 - 4000 pmol/L (0.302 - 12.1 µg/L)


Feature

ü  No or minimal cross-reactivity to human insulin and proinsulin

ü  High specificity enables measurement of human C-peptide in mouse and rat models

ü  CE/IVD labeled on selected markets

ü  Serum controls available: 10-1134-01 (US only) or 10-1164-01

 

Description

The C-peptide ELISA is a high quality immunoassay for specific quantification of human C-peptide in serum, plasma or urine

The Mercodia C-peptide ELISA is a highly specific tool registered for in vitro diagnostics (IVD) use. It is calibrated against the International Reference Reagent for human C-peptide, IRR84/510. The specificity of the utilized antibodies enables measurement of human C-peptide without interference from insulin or proinsulin, and measurement of human C-peptide in mouse and rat models.

Each Mercodia C-peptide ELISA kit contains reagents for 96 wells, sufficient for 42 samples and one calibrator curve in duplicate. The microplate strip format (12 strips x 8 wells) is ideal when only a limited number of samples need to be analyzed.

Mercodia Diabetes Antigen Control (10-1134-01(US only) or 10-1164-0) is designed to be used as a two-level control for human C-peptide.

 

Test characteristics

Species  Human

Samples  Serum, EDTA and heparin plasma or urine

Sample volume  25 µL

Assay range  100-4000 pmol/L (0.3-12.0 ng/mL)

Detection limit  ≤ 25 pmol/L (0.076 μg/L)

Incubation  1h + 1h + 15 min

Detection method  Absorbance

Specificity

The following cross-reactions have been tested:

Insulin  < 0.0006 %

Proinsulin  2 %

Proinsulin (Des31-32)  3 %

Proinsulin (Split 32-33)  2 %

Proinsulin (Des 64-65)  74 %

Proinsulin (Split 65-66)  10 %